TO: Ken, KC6TEU, and the Microwave Group 04/27/02 From: Dick, K2RIW RE: Directional Couplers, and a VSWR/Power Measurement Procedures.
DIRECTIONAL COUPLERS USED FOR VSWR AND POWER MEASUREMENT INTRODUCTION -- Over the years I have heard many engineers, and some smart amateurs, express opinions that reflect a considerable misunderstanding about the operation of Directional Couplers, and how to properly use them in the measurement of Voltage Standing Wave Ratio (VSWR), and power. This memo is intended to give some basic information that may help.
At first, the average electronic technologist is mystified by at least two
of the concepts of how RF behaves within transmission line structures:
I believe that both of these principles must be absorbed (and understood),
if meaningful DC measurements are to be properly executed, and believed.
Here are my recommended procedures, with some partial explanations of what
is taking place at each step.
A DIRECTIONAL COUPLER USED IN A VSWR OR POWER MEASUREMENT PROCEDURE
(I) DIRECTIONAL COUPLER CALIBRATION -- The first step in this procedure is
to establish the quality of the Directional Coupler (DC) that you are about
to use. I don't care if the label on the DC says it is a "Cadillac" or
"Rolls Royce" brand, and the calibration sticker says it is traceable to
"The Bureau of Standards" with and accuracy of 0.01 dB; you still have to
confirm that it is good working order right NOW. It is possible that the
DC was thrown onto a concrete floor yesterday, and the internal termination
may have been shattered. If that had happened, it could loose almost all
of it's directional characteristics -- it's "Directivity."
The confirmation requirement is similar to the proper use of an Ohm Meter.
Notice that a good technologist will always short the two leads together;
and the Ohm meter had better read a small fraction of an ohm, before the
technologist will proceed with the next measurement.
Similarly, a prudent technologist will measure the Directivity of the DC
he is about to use. It is also useful to know that sometimes the DC can be
used far outside the frequency range it was designed for, as long as the
principle of operation is somewhat understood, and a calibration at the
present frequency is performed. Here is the Directivity Confirmation
procedure.
DIRECTIVITY CONFIRMATION -- Unfortunately, the Directivity Confirmation
procedure requires a known good termination (dummy load), and the procedure
will have an accuracy that rarely is much better than the quality of that
termination being used. First apply an RF signal to the DC "input" port,
with a known good termination connected to the "output" port. Position the
DC so that it favors the Forward flowing signal. Place a power-measuring
device at the directional port. This can be a Power Meter, Spectrum
Analyzer, calibrated Crystal Detector, Scalar Network Analyzer, etc.
Measure (and record) the DC's response to the forward-flowing signal (in
dBm units). If, for instance, you are using a known Directional Coupler
(DC) with a -10 dB Coupling Coefficient, the measured power should be
nearly 10 dB weaker that the power that's being applied from the signal
generator. By the way, "dBm" means Decibels of signal strength with
reference to a 1 milliwatt signal.
Next, reverse the DC "input" and "output" ports, and repeat (and record)
the previous measurement. The difference in the two readings indicates the
Directivity. For instance; if a 0.0 dBm signal generator is applied to a
10 dB coupler, and it measured -10 dBm during the Forward Measurement, and
-30 dBm during the Reverse measurement, that would indicate a Directivity
of 20 dB (the difference in the readings). A DC of "Good" quality will
show a directivity of 20 dB, that is, the apparent reflection from the
termination will appear to be -20 db (an apparent VSWR of 1.22:1), even if
the termination is a perfect 50 ohm resistance at the present frequency.
An "Excellent" DC will show a Directivity of 30 dB (an apparent VSWR of
1.065:1), and there are Instrumentation-type DC's that can display a
Directivity of over 50 dB (an apparent VSWR of 1.006:1). More on this
later; there are ways of improving your DC's Directivity.
Simplistically, you could say that a DC that displays a Directivity of 20
dB will not be able to easily resolve the Reflection Coefficient from an
unknown load of better than about -20 dB, there are ways to get around
this. Depending on how well your DC is internally balanced, the finite
Directivity (-20 dB for instance) represents the degree of response it has
to a signal that is flowing in the wrong direction -- this is really it's
degree of imbalance. A modern Network Analyzer uses a complicated "12
point" calibration procedure to drastically improve the accuracy of a
Reflection measurement it makes with it's "only Good quality" Directional
Couplers.
ALTERNATE CALIBRATION PROCEDURE -- There is an alternate Calibration
Procedure that does not require the inconvenience of reversing the DC to
measure it's Directivity. This is to recognize that a good Short (or Open)
circuit has a Reflection Coefficient of nearly -0.0 dB. In this method,
first measure (and record) the apparent reflected power from a Short (or
Open) termination, then place the Known Good Termination on the "output"
port of the DC and repeat the measurement. The difference (in dB) between
the two measurements represents the DC's Directivity. When using SMA or
type N connectors at 10 GHz (and below), an "Open Circuit" will have
Reflection Coefficient of nearly -0.0 dB, and is a good calibration
"short/open termination." However, if you're using a Wave Guide (WG) type
DC, an open circuited WG flange makes a pretty good transmitting antenna,
with a VSWR of about 1.5:1 (reflection coefficient of about -12.9 dB).
Therefore, don't use this as a high reflection termination. Instead, place
a sheet of metal (tightly) across the WG flange as the high reflection
termination.
SIGNAL GENERATOR VSWR -- There is an additional danger to the alternate
calibration procedure. It is vulnerable to the VSWR of the signal
generator. I would only use this procedure if there was a 10 dB (or
greater) pad between the signal generator and the DC. Without that pad,
the reflected signal could re-reflect from the signal generator and cause a
confusing reading. The signal-generator-reflected voltage can add to the
incident voltage and create an apparent signal source that would appear as
much as 6 dB greater (or more) in magnitude -- but only during the
short/open portion of the test. Also, if the DUT happens to have a rather
high VSWR (reflection of greater than say -20 dB), I again would recommend
the use of a 10 dB pad at the signal generator.
(II) THE UNKNOWN MEASUREMENT -- Once you have confirmed that your DC is
performing properly, it is time to place the Unknown Circuit (the Device
Under Test [DUT] ) on your DC to measure, and tune, it's Reflection
Coefficient. The DUT-reflected signal can then be translated into VSWR by
using a look-up table or by performing a two step calculation. Step (1):
Convert the reflection coefficient (in dB) into a reflection Voltage, which
is usually represented by the Greek letter Rho. Step (2): Convert the Rho
magnitude into VSWR.
The final dB of Reflection Coefficient in the numerator must be a negative
number that's then divided by 20 and raised to the power of ten in formula
(1). At first, some technologists will understand that the dB value is
negative dB's, they place it into the formula that has another negative
sign in it, that converts it to a positive value (+), and they come up with
answers that are crazy.
CHEAP AND BROAD -- The beauty of using a Directional Coupler (DC) in VSWR
measurement is that, generally, they are rather inexpensive, and they are
rather broadband, therefore a swept frequency measurement is possible if
your power detector is a fast acting one, such as a calibrated Crystal
Detector (and oscilloscope), a Spectrum Analyzer, or a Scalar Network
Analyzer (SNA). As you tune your DUT, it is nice to know that you are
tuning for a broadband match, as apposed to an impedance match that is only
effective across a narrow frequency range.
(III) DC ALTERNATES -- There are a large number of devices that can serve
as the Directional Coupler (DC). They have such names as Quadrature
Hybrid, 90 Degree Hybrid, Branch Hybrid, Branch Coupler, Magic T, Ring
Hybrid, Zero-180 Degree Hybrid, Wave Guide Broad Wall Coupler, Wave Guide
Narrow Wall Coupler, Wave Guide Beth Hole Coupler, etc. The one kind of
hybrid that can't be used this way is a Wilkinson Half Hybrid, or Zero
Degree Hybrid.
(IV) DC EXTENDED FREQUENCY RANGE -- Few technologists know that a well
constructed Directional Coupler (DC) has an operational frequency range
that extends many octaves in the lower-frequency direction. For instance,
if you plotted the Forward Response of a DC that's rated for operation from
1 to 2 GHz, you would find that it has useful operation all the way down to
10 MHz (and probably below). The only thing that changes is it's frequency
flatness, and the Coupling Coefficient decreases -- but that can ba a
considerable advantage. Here is what's happening:
(A) A TEM-type (non Wave Guide type) Directional Coupler has it's greatest
coupling at the frequency where the internal coupling section is 1/4 wave
long. Above (and below) that frequency the response falls off in a very
predictable manner -- it's a SINE wave of amplitude. In other words, if I
was sweeping that DC that's rated for 1 to 2 GHz, and I plotted the
Foreword absolute Voltage response versus frequency at the Coupled Port,
the resultant plot would look like a rectified SINE wave, with the
horizontal axis being frequency (instead of time). There would be a zero
response a zero MHz, a broad peak near 1.5 GHz, a second zero near 3 GHz, a
second broad peak near 4.5 GHz, etc. Unfortunately, a DC only has
Directivity at the 1/4 wavelength frequency region, and at lower
frequencies -- but that still leaves many octaves of useful operation.
(B) That predictable response outside of the rated frequency range has turned into an advantage for me on many occasions, here are some examples:
(1) For my first published article, "A Stripline Amplifier/Tripler for
144 and 432 MHz", Ham Radio, February, 1970, I needed to test the power
output, and harmonic content, of the 144 MHz section and the 432 MHz
tripler section of that 4CX250B amplifier. I needed a 300 watt
frequency-indicating power meter, that I didn't have. A Spectrum Analyzer
(SA) can do the job, but it can't tolerate the 300 watts. If I had a -30
dB DC, the coupled power would be 0.3 watts and the SA could easily make
the measurements. But, my company's Instrumentation Department said they
didn't have a -30 dB DC at that frequency range, and non of their DC's
could tolerate 300 watts.
I studied what they had and found a solution. They had a Narda -10 dB
type-N Directional Coupler rated for 8 to 12 GHz and 1 watt maximum. I
reasoned that the coupling section was 1/4 wave long (90 degrees in phase
length) at 10 GHz, the center of it's frequency range. I then divided 144
MHz by 10 GHz, multiplied by 90 degrees, and reasoned that the coupling
section was only 1.296 degrees long at 144 MHz. The SIN of 1.296 degrees
is 0.02262. Since this is a voltage response I took 20*LOG(0.02262) =
-32.9 dB. That means that the coupled response at 144 MHz would be -32.9
dB (weaker) than at 10 GHz, where it was a -10 dB coupler. Therefore it is
a -42.9 dB coupler at 144 MHz. I calibrated it at 144 MHz and found it to
be a -43.1 dB coupler -- close enough. And, since the internal coupled
line is isolated from the main line by -43.1 dB, that means that the
internal 50 ohm termination would never see more than 0.015 watts when I
applied 300 watts of 144 MHz signal to the coupler. I similarly calibrated
it at the harmonic frequencies, applied the 300 watts to it, it worked like
a charm, I made all the measurements this way, and they appeared in the
article.
(2) In the low frequency area of a coupler's response (near 0 degrees of
a SIN function) the response is almost a straight-line response that falls
off at -6 dB per octave (-20 dB per decade) as you go down in frequency.
Therefor the "-43.1 dB coupler" I used at 144 MHz would be a -63.1 dB
coupler at 14.4 MHz. As you are about to see, Directional Watt Meters use
this principle.
(V) BIRD-TYPE WATT METERS -- It is interesting to note that the slug of a
Bird Watt Meter is also a less than 1/4 wave section of a Directional
Coupler. The Bird slug achieves frequency flatness across it's rated
frequency range by using a rectifier circuit that has a low-pass filter
action that rises at 6 dB per octave as you go down in frequency.
Each slug also has a finite Directivity, depending on how well it was
balanced and calibrated at your favorite frequency. Therefore, be careful
about falling into the trap of using a high power slug to measure the
forward power of your 1 kw XMTR, and then switching to a low power slug to
measure a very low VSWR. Your antenna may be perfect, and have no
reflected power (voltage), but the slugs approximate 20 dB of Directivity
would show an apparent antenna reflection of -20 dB (10 watts). That would
lead you into believing that the antenna VSWR was 1.22:1.
(VI) COUPLER IMPROVEMENT TECHNIQUES -- AS the above material shows, a DC
that has less than ideal Directivity is really displaying a slight
imbalance that causes it to slightly respond to the signal that is flowing
in the wrong direction on the main line of the coupler. There are many
ways of improving the DC's balance.
(1) Internally, you could re-adjust the accuracy of it's termination, or
you could add a small gimmick capacitor in the correct location to improve
the Directivity balance.
(2) But, an even better way is to use a Double Slug Tuner, or a Wave
Guide E-H Tuner. If you have a known good termination, you can assume that
it has perfect absorbtion and essentially no reflection. You then place
the tuner between the DC and the good termination, and adjust it until the
DC shows no reflected power from the termination. You then leave the tuner
connected to the same port of the DC, while you proceed with the VSWR or
power measurements. When you were adjusting the tuner for a null in the
DC's Reflection response, you were really creating a second small reflected
signal that was equal in amplitude and 180 degrees out of phase at the DC
coupled port. That created the improved balance and made the DC nearly
ideal, at that frequency. The bandwidth of this DC correction technique is
dependent on the amount of correction that was required. When in doubt,
recheck the balance at the next frequency.
(VII) TRANSMISSION LINE DIRECTIONALITY -- When I tell a technologist that
a transmission line will keep the two signals completely separate, that
flow in opposite directions on a transmission line, they often don't
believe it -- particularly if the two signals came from the same source.
There are many RF tests that could be performed to prove this, but I have
discovered that a well-informed sceptical person can always come up with an
alternate explanation that supports their point of view. I have found that
the best way is to use visual experiments.
(1) A pool of water is really a radial transmission medium. If I drop a
pebble at the North end of the pool, waves will travel to the South.
Similarly, a pebble dropped into the South end will create waves that
travel to the North. If I drop pebbles at both ends of the pool, the waves
will meet at the middle, and pass right through each other with no
interference, as long as the waves are kept small enough (use the linear
region of wave amplitude -- no white caps).
(2) I can tap the 1/4 inch guy wire on my 200 foot Rohn-55 tower and watch
the wave travel up the guy wire, strike the tower, reverse in polarity, and
propagate back down to me (it hit a "short circuit"). I can wait until the
wave has struck the tower, and started back to me, then I can strike the
wire again (with any polarity) to start a second wave going up the guy
wire. As the two waves meet in the center, they pass right through each
other with no interference, as long as the waves are small enough that I
don't get into non-linear stretch (deflection) of the steel.
(3) I say that most linear transmission mediums obey this property -- even
RF in free space. Those waves that meet in free space pass through each
other with no real interference. When you move your Handy Talky Radio
around a room that is reflective, you will find what you think are signal
nulls. This is because you are using an antenna that has no Directivity,
and it is responding to at least two waves that are out of phase.
Similarly, the probe that is used on a Slotted Line VSWR setup has no
directivity, and it displays the Standing Wave Ratio that is caused by the
signals that flow in both directions through the Slotted Line. This
measurement technique has become the classic way of specifying the
Reflection Coefficient of an RF device -- it's VSWR.
(VIII) LETS DO AWAY WITH VSWR -- If you took the directional probe from
the slug of a Bird Watt Meter and operated it on that Slotted Line, you
would discover that the Standing Wave has disappeared, and you could now
independently measure the amount of power (or voltage) that is flowing in
each direction (by reversing the slug) -- that's really what you wanted to
know in the first place.
In the past, that Slotted Line measurement was the only way you could
conveniently measure the reflected voltage -- by using an interferometry
technique to indirectly measure it as VSWR. It really is time that we
abandon "VSWR measurements" because we don't do it that way any more. We
should only discuss the Reflection Coefficient -- in watts ratio, volts
ratio or dB ratio (choose your favorite units), because we now directly
measure the reflected signal. We RF mavens seem to spend half our life
converting back and forth between VSWR, Voltage Reflection Coefficient
(S11, S22) or Power Reflection Coefficient, just so that we can communicate
with a technologist (or the data sheet) that uses the other system of units.
"VSWR" is now a "coded message," it's really time that we "Break the Code"
or stop using that code when we're training the new RF recruits. I'll
admit that we will have to keep mentioning it, for historic reasons.
(IX) TROMBONE IMPROVEMENT -- I'll warn you that these last three paragraph
will only be appreciated by a person with a rather exacting-type of
personality.
Once you accept the fact that RF power can independently flow in two
directions on a transmission line, you then realize that changing the
length of a lossless transmission line does not change the Reflection
Coefficient; thus it doesn't change the true VSWR of your antenna.
However, if the Directional Coupler (DC) device your using (coupler or a
Bird) has less than ideal Directivity, than the Reflection Coefficient, and
VSWR, will appear to change. This is because there is a small amount of
Forward-flowing signal (I'll call it the Leakage Signal) that's mistakenly
being picked up by your coupling device, that beats against the real
Reflected Signal that your coupler is now measuring (from your antenna, for
instance). As you change the length of the transmission line (with a
Trombone Line), the two signals go in and out of phase with each other.
This will show up as a cyclicity of the apparent Reflected Signal Power, as
the Trombone is operated. This assumes that your trombone can move about
one wavelength at your frequency -- you not going to do this at 80 meters,
Hi. Although, there you could insert fixed lengths of low loss cable to
get the same effect.
Knowing the operation of the system, and its shortcomings can allow you to
gain a higher accuracy in the Reflection Coefficient measurement. A
perfect DC or Bird would show no change in reading as the Trombone (on the
antenna side) is operated. The magnitude of the "ripple" is an
interferometry effect that is telling you exactly how strong is the Leakage
Signal into your coupling device. Once you know the strength of the
Leakage, you can subtract it out of your measurement. This is exactly the
accuracy improvement procedure that is done in the microprocessor of a
modern Network Analyzer. You can convert the Ripple into a Leakage
Magnitude by using the following formulae:
Here is a measurement example. Assume I'm measuring the Reflection Coefficient of my UHF antenna system and my DC says that the Reflection is around -19.5 dB. As I operate the Trombone after the Coupler, I see a Peak reading of -19 dB, and a valley reading of -21 dB. That's a Peak-to-Peak reading of 2 dB. The formula tells me that my Leakage Signal is 0.1146, or -18.81 dB (weaker) than the Peak and Valley measurements I have made. That relative Leakage voltage was in-phase at the -19 dB reading, and out-of-phase at the -21 db reading. I can choose to subtract the voltage from the -19 dB, or add it to the -21 dB reading. This relative voltage will thus be 1.1146, or 0.9954 (as a voltage), and I can take 20*LOG of these voltages. Thus, I can either add 0.94 dB (in absolute terms) to the -19 reading, or subtract 1.06 dB (in absolute terms) from the -21 dB reading. In either case the corrected reading will be an antenna Reflection Coefficient of -19.94 dB. I hope this information is useful to those who could read this far. Feel free to correct the mistakes.